TYPE
|
Insulation resistance meter
|
PHYSICAL SPECIFICATIONS
|
19”W x 7”H x 15”D, (48 cm x 17 cm x 38 cm); Weight: 24lbs (10.9 kg)
|
INPUT POWER
|
110 – 120 Vac or 220 – 240 Vac (selectable), 50/60Hz
|
RESISTANCE READING RANGE
(0°C to +30°C)
|
100 K-ohms – 1 M-ohm ±20%,
1 M-ohm – 1 T-ohm ±5%,
1 T-ohm – 5 T-ohm ±20% , 5kV
100 K-ohms – 1 M-ohm ±20%,
1 M-ohm – 100 G-ohms ±5%,
100 G-ohms – 500 G-ohms ±5%, 500 V
1 M-ohm – 10 G-ohms ±5%, 50 V
|
TEST VOLTAGE
|
Selectable from 50Vdc – 5 KVdc, in 2 Vdc steps
|
OUTPUT VOLTAGE ACCURACY
(0°C to +30°C)
|
±2%, ±1v of selected voltage with load resistance greater than 100 MΩ
|
SHORT CIRCUIT CURRENT
|
2 mA max
|
CURRENT READING RANGE
(0°C to +30°C)
|
0.03 nA – 2 mA; Accuracy: ±5% ±0.2 nA
|
CAPACITANCE READING RANGE
(0°C to +30°C)
|
0.01 μF – 10.0 μF (Test voltage greater than 100V); Accuracy: ±5% ±0.03 μF
|
VOLTAGE READING RANGE
(0°C to +30°C)
|
50 – 1250 Vac (rms) or dc; Accuracy: ±5%, ±1V
|
CAPACITOR DISCHARGE
|
Less than 2 Sec/μF, automatic at the end of test
|
HUM REJECTION
|
1mA per 1kv of test voltage, 2mA rms maximum
|
BATTERIES
|
Two 12V, 2.0 Ah sealed lead acid batteries. Battery life: typical 6 hrs, continuous testing. The IRM-5000P can be used while charging.
|
PRINTER
|
Built-in 2.5-inch wide thermal printer can print test results in both tabulated and graphic formats
|
COMPUTER INTERFACE
|
One RS-232C port
|
SAFETY
|
UL Certified (UL 61010A-1), CAN/CSA Certified (C22.2 No. 1010.1-92)
|
ENVIRONMENT
|
Operating: -10? to 50? C (15? to +122? F); Storage: -30? C to 70? C (-22? to +158? F)
|
CABLES
|
One 6-foot cable set, one 15-foot cable set, ground cable, power cord, cable bag
|
OPTIONS
|
Transportation Case
|
WARRANTY
|
One year on parts and labor
|
Specifications are valid at nominal voltage and ambient temperature of +25?C (+77?F).
|